FRIWO - FOX12-XM-USB Medical Power Supply 5V / 2200mA worldwide
-
High efficiency
-
Low no load power consumption 50mW
-
Meets DoE efficiency level VI
-
Meets CoC version 5 Tier 2 ready
-
Ultra-low leakage current ≤10μA
-
IEC 60601-1-2, 4th edition ready
-
Wide operating temperature range 0 - 45°C
-
Isolation 2x MOPP
-
Double housing safety: snapped & welded
-
Also available as version with fixed plugs: FOX12-FM-USB
Required Accessories
For the operation of this device you need a primary adapter. This can be found here: Primary adapters
Product information "FOX12-XM-USB"
IEC 60601-1 approved, ≤ 10μA leakage current, Protection class 2xMOPP, DOE 2016 ready, IP42 approved exchangeable plug system as additional accessories, Double Safety: Housing snapped & welded.
Additional information
- Input voltage: 100-240 V
- Input tolerance: ±10%
- Frequency: 50-60 Hz
- Input current: 300-150 mA
- leakage current: ≤10 µA
- Overload protection: Yes
- Over Voltage Protection: Yes
- Short Ciruit protection: Yes
- Working temperature: 0°C - 45°C
- Humidity: 10-95% V
-
Storage temperature
: -40-70°C - Operating Altitude: 3.000 m
- Means of Protection: Input Output 2xMOPP, Input - Ground 1xMOPP
- Protection Class: II
-
MTBF: 200.000h acc. Mil217F (based on calculations at
120Vac/60Hz & 230Vac/50Hz, ambient 25°C
and 100% load) (The MTBF are theoretically determined values, which does not guarantee the lifetime of the product or of the electrolytic ) h - AC Input: FRIWO interchangeable mains plug system: see Primary Adapters
- DC output: USB socket Typ A
- Gehäuse / Housing: FOX12-USB
- material: PC-ABS
- Output voltage tolerance: ± 5%
- Spec.0 cover sheet FW-Type: FW8000MUSB/05
- Dimension: 77,5x31,5x41
- Standard: IEC60601-1
- Approbations: Europe, USA, Australia, Japan
- Housing Flame Class: UL 94 V0
- IP Rotection rating: IP40
- Line Regulation: ±1
- Load Regulation: ±4
- Atmospheric pressure: 70-106
- Storage Humidity: 10-95%
- Hold up Zeit: >10ms(120V), >50ms (230V)
- Inrush current: <35
- Condcted and Radiated Emmisions: EN55022, EN55011, FCC15, Class B, EN60601-1-2, IEC60601-1-2 4th Edition
-
Immunity
: EN55024, EN60601-1-2, IEC60601-1-2 4th Edition - Harmonics: EN61000-3-2
- Flicker noise: EN61000-3-3
- ESD (contact/ air): EN61000-4-2
- Immunity against radiated field: EN61000-4-3
- ETF/ Burst: EN61000-4-4
- Surge: EN61000-4-5
- Immunity against conducted disturbances: EN61000-4-6
- Voltage dips: EN61000-4-11
- Immunity against magnetic field: EN61000-4-8
- Vibration: Standard: IEC 60068-2-6, Frequency range : 10 – 500 Hz, Cross- over-frequency : 58 – 62 Hz, Displacement amplitude : 0.2mm, Peak acceleration : 3 g, Number of cycles: 10 per axis